The HRMA/ACIS effective area and energy resolution were also evaluated for split event thresholds of 10 to 16 adu in steps of 1 adu.
Figure 6.5 and figure 6.6 show the dependence of effective area with split event threshold for ACIS CCD's I1 and S3. Tabular versions of the data are listed in Tables 6.12 and 6.13. Figure 6.7 and figure 6.8 show the dependence of energy resolution with split event threshold for ACIS CCD's I1 and S3. Tabular versions of the data are listed in Tables 6.14 and 6.15.
These preliminary results indicate that the HRMA/ACIS effective area and energy resolution do not vary considerably with split event threshold for FI CCD's, whereas the BI chip performance is significantly different, showing a significant dependence on split event threshold.