Next: Calibration Strategy
Up: ACIS UV/Optical Blocking Filter
Previous: ACIS UV/Optical Blocking Filter
The X-ray ring at NSLS in normal operation mode accelerates electrons to
energies of about 2.5 GeV for optimized X-ray emission in the range of 1
to 10 keV.
A table of the operating parameters of the NSLS X-ray storage ring is
provided at their Web site (http://www.nsls.bnl.gov/). Our experiment
was set up on beamline X8A, one of 56 X-ray beamlines attached to the
X-ray storage ring.
The beam energy incident on the filters was selected by adjusting the
orientation of the crystal element of a double crystal monochromator located upstream
from the vacuum chamber. Each crystal element spans a certain energy range.
In particular a W/Si multilayer
covers the energy range 0.26-2.0 keV with energy resolution
(
) of 2x10-2
a Si(111) crystal covers the range 2.1-5.9 keV
with a resolution of 5x10-4 and a Beryl(1010) crystal covers the
range from 0.8-2.0 keV with a resolution of 8x10-4.
In Table 5.1, we list the
energies surveyed and the monochromator crystal elements used
for the transmission measurements of the ACIS-I and ACIS-S
UV/Optical blocking filters. Detailed fine energy scans were performed
above the absorption K edges of C, N, O, and Al to investigate extended X-ray
absorption fine structure (EXAFS) and shifts in the location of the Al-K edge due
to possible Al2O3 contamination.
The configuration of the test set up is shown in Figure 5.1. The ACIS filters were placed into the UC/SAO spectrometer-reflectometer chamber and mounted onto a mechanical fixture. The fixture was made to support both image and spectrometer filters and was mounted on a rotary stage that allowed for the selection of the portion of the filter to be intercepted by the X-ray beam. The X-ray beam was collimated with a set of vertical and horizontal entrance slits to a size of 2 mm by 2 mm. The beam flux was monitored using a detector that was moved in and out of the X-ray beam with a computer controlled actuator.
![]() |
Mark Bautz