Go to HETG at XRCF
The results to date suggest adopting the following calibration values:
Grating | dp/p rms | error | arc min rms | error |
---|---|---|---|---|
MEG | 162. | 50. | 1.5 | 0.5 |
HEG | 146. | 50. | 1.5 | 0.5 |
For the period variations I have used the sub-assembly value rss'ed with 100 ppm as an estimate, with an error of +/- 50ppm.
For the roll variations I have used 1.5 arc minute rms with an error of 0.5 arc minute. This is based on the MEG HSI FC data set described below which gives a measured MEG roll rms less than 2.07 arc minute (other effects have not been removed so this is an upper limit.)
LR sub-assembly data predicts the values in the table below for dp/p rms and the grating roll rms values.
Grating | dp/p rms | error | arc min rms | error |
---|---|---|---|---|
MEG | 127. ppm | 10. ppm | 0.42 | 0.1 |
HEG | 106. ppm | 10. ppm | 0.42 | 0.1 |
The period and dp/p values measured with the LR system should be accurate - the main additional effect in going to the full flight model are any period variation effects due to the mounting process (aligning and screwing the frame to the HESS).
The existence of the "mis-aligned" MEG gratings shows that the polyimide can introduce polarization effects that can cause the X-ray roll angle to differ from the polarization-measured roll angle. The subassembly roll values are the polarization measured ones. It is possible that the same phenomenon is working to a smaller degree for all the MEGs and HEGs: polyimide polarization effects will lead to X-ray roll variations greater than the sub-assembly predicts.
Grating | Energy | Order | E/dE | dp/p rms limit |
---|---|---|---|---|
HEG | 1.254 | 1 | 1049. | < 405. ppm |
HEG | 1.254 | 2 | 1418. | < 300. ppm |
MEG | 1.254 | 1 | 479. | < 888. ppm |
MEG | 1.254 | 3 | 1095. | < 389. ppm |