ACIS Instrument Support - Monthly Progress Report for October 2010 1. ACIS Characterization 1.1 Characterize and model radiation damage in ACIS CCDs ... Continued work on irradiation/annealing tests done at GSFC in 2005. Writing up as a paper. 1.3 Monitor scientific on-orbit performance of ACIS ... Continued monitoring of ACIS CTI, gain, trailing pixel pulseheight, detector efficiency, corner pixel distribution, spectral line centroid and width, and focal plane temperature. Updated monitoring web pages. (http://space.mit.edu/~cgrant/monitor.html) ... Continued to recalibrate CTI-sacrificial charge correction used in monitoring. Finalizing new correction factor. ... Examining further how shorter exposure times (< 5 ksec), yield biased CTI measurements. Ongoing. ... Checked for any performance changes before and after the recent Crab observations. None seen. 2. Engineering 2.1 Instrument hardware engineering performance ... Continue to monitor long term ACIS instrument performance. All ACIS engineering parameters continue to operate within predicted limits. 2.3 Instrument science performance ... Continued investigation into whether ACIS can act as it's own radiation monitor, prompted by action from the thermal working group which is concerned about the continued degradation of EPHIN. Further investigation of the false trigger frequency and false trigger source characteristics. ... Participated in monthly ACIS operations meeting. Discussed modeling of ACIS temperatures and using ACIS as a radiation monitor. ... Participated in radiation working group telecon. ... Continued daily quick-look processing of ACIS telemetry and maintained Web interface to ACIS real-time data. ... Continued to publish the "annotated DOT" timeline reports. 3. CXC Support 3.1 Provide user support ... Updated ACIS memo listing calibration ObsIDs and observation characteristics (ACIS Memo #165). 3.2 Provide data analysis support ... Participated in CXC Cal/Ops modeling discussion group. Discussed release of temperature-dependent CTI correction software (!), CC-mode calibration and the status of contamination calibration. ... Supplied CXC with dates and times for the data with mis-calibrated temperatures from September 2005. Confirmed that their strategy for the temperature-dependent CTI correction in this time period was sound. ... Continued to maintain ACIS calibration database of ground and flight data products. 4. OCC Support 4.1 Use of MIT/ACIS EGSE at OCC ... Continued support and maintenance of ACIS EGSE software and hardware at the OCC. Outreach: None