ACIS Instrument Support - Monthly Progress Report for April 2003 1. ACIS Characterization 1.1 Characterize and model radiation damage in ACIS CCDs ... Continued running charge injection tests on S2 from the command load. Two tests repeating earlier runs with warm focal plane temperatures. [... Preliminary analysis of charge injection tests shows a linear dependence of CTI measured from cal source lines and measured from charge injection. Working on better determining measurement errors. [... Preparing new set of charge injection tests on I3. 1.2 Monitor scientific on-orbit performance of ACIS ... Continued analysis of ExtCalSrc data to search for efficiency decrease at low energies. Worked on improving analysis of 1.5 keV line, to reduce systematic errors. ... Continued monitoring of ACIS CTI. No significant changes beyond the long term slow increase. (http://space.mit.edu/~cgrant/cti/) ... Continued monitoring of ACIS gain. All CCDs stable with a slow decline in gain slope (ADU/eV) except for I2 which continues to vary. (http://space.mit.edu/~cgrant/gain/). 1.4 Update and improve CCD simulation ... Continued work on improving FI and BI CCD models with a more physical representation of the electron traps that cause CTI. 2. Engineering 2.1 Instrument hardware engineering performance ... Continue to monitor long term ACIS instrument performance. A review of the past 60 days of engineering data reveals no new trends or anomalous behaviour within ACIS. ... Continue to support thermal monitoring of the PSMC. Ongoing monitoring of thermal conditions on the SIM has resulted in recommending a new set of thermal warning limits for the instrument. [...] Reviewed DEA housekeeping temperatures. No large deviations. Based on previous history, new warning limits are being determined. 2.2 Flight software maintainence [...] The ACIS engineering team released Rev C of the Flight Software Optional Patches. This release added two new diagnostic timed-exposure modes (patches ctireport1 and ctireport2), a fix to a hardware race condition (patch untricklebias), and a fix to a logical bug in raw data compression (patch compressall). 2.3 Instrument science performance [...] Finalized new CC mode bias parameters. New parameters will be tested on board during a CTI run and then implemented for all CC mode science observations. ... Continued daily quick-look processing of ACIS telemetry and maintained Web interface to ACIS real-time data. ... Continued to publish the "annotated DOT" timeline reports. [...] Participated in ACIS telecon. [...] Participated in ACIS Operations meeting. ... Continued to support charge injection operations by reviewing loads and comparing flight data to ground data to select new settings. [...] Participated in Chandra quarterly review. Presented ACIS status report. 3. CXC Support 3.1 Provide user support ... Updated ACIS memo listing calibration ObsIDs and observation characteristics (ACIS Memo #165). 3.2 Provide data analysis support [...] Participated in CXC Cal/Ops RMF discussion group. [...] Checked magnitude of pulseheight evolution from CTI and instrumental gain changes for I2. Instrumental gain change dominates! ... Continued to maintain ACIS calibration database of ground and flight data products 4. OCC Support 4.1 Use of MIT/ACIS EGSE at OCC ... Continued support and maintenance of ACIS EGSE software and hardware at the OCC. all recent memos available at http://space.mit.edu/ACIS/iacis.html Outreach: none