Anomalies seen in the Plots created by rad_compare updated 5/7/01 dd ------------------- Key: ** - anomaly not yet explained or investigated OK - anomaly is explained !! - anomaly is or could be a "real" measurement change :) - no unexplained anomalies ------------------- MA1025 (010301/001207) OK - zero jump at 0.93 keV (high zo Chi2) 4/4-OK: SSD/fitting problem, reanalyzed, lowered Chi2 OK - first second-order point consistantly ~12% low in all regions 5/8-OK: fits look good. Error on ref ratio is 6% one sigma - ref stat error could create this easily. Note that the Chi2 is not large for the regions. :) MA1026 (010301/001101) :) MB1148 (010302/001117) !! - zero-order change at MB1148m4 +15% at 0.93 keV 4/4 data look alright, real change or just stat fluke? OK - first-order 0.93 keV 10% high in m4 and all regions !! 5/4 Conclude that it's probably either: i) some ref/PSPC thing common to '1148 and '1158 ( and also showing up in some HEG tests...) OR ii) it may be a rad effect that BOTH '1148 and '1158 experienced, if they were both rad-exposed. MB1158 (010302/001117) OK - second order Chi2 large - poor fitting of small signals 4/4 exclude poor fits/measurements :) MA1022 (010307/001116) :) ME1409 (010307/001127) OK - large second-order Chi2 in m=-2 Mg in region p8 5/8 low number of second order counts gave rise to a poor fit (115 counts in ROI yet gaussian fit gave 134...) :) HC2214 (010308/001101) OK - large second-order Chi2 5/7 3 regions OK, 2 regions CHi2 < 6.0. :) HC2259 (010308/001101) OK - minus 20% change in first-order 0.93 keV m=+1 effic in p8, !! other regions too 5/4 improved HEG Cu-L line fitting and effect is more like only -10% averaged over the five regions. Unfortunately we don't have the 001101 m=+1 Cu-L measurements for HC2214 which was tested at the same time as '2259... This effect feels like the MB1148 problem of Cu-L measurement systematic effects... 5/7 The HX220 ratios are within 3% for 001101 and 010308, so most of the 10% effect is from the grating... and counting uncertainty of few per cent. OK - second-order large change in region m8 at one energy, 1.49 keV 5/8 A complete mis-fit of the 2nd order 010308 distribution: 634 ROI counts yet Gaussian fit is only 167. HB2104 (010309/001113) ** - first-order in p4 region looks 5-10% low over all energies -- this is isolated to one test region of the grating -- don't see anything like this in other gratings/regions, what's the cause? Same effect is reducing the first two second order points (1.25 and 1.49 keV) in p4? ** - large zero-order Chi2, 4 of 5 regions have 5% drop in effic at 1.25 keV ** - large second-order Chi2 in region p0 at m=-2 4.5 and 6.4 keV HE2426 (010309/001115) OK - large second-order Chi2 in m8 (Al-K +1 order) 5/7 re-analysis reduced the Chi2. :) HE2412 (010312/001115) OK - large zero-order Chi2: 5% drop in effic at 1.25 keV 5/7 OK: looks statistical... :) HF2517 (010312/001114) OK - first-orders at 0.93 keV: one higher, one lower by ~15%. !! 5/4 re-analysis reduced the effect a little... a la '2259. 5/7 The reference measurements are only 4% different, so it appears to be mostly in the grating... ** - large zero-order Chi2: ~5% increase in effic at 1.25 keV ** - second-order Chi2 at one energy in m8 region ------------------------------------------------------------------